SM57Q65 automatic filtration, sampling, and cleaning system
-
SM57Q65 Automatic Filtration, Sampling, and Cleaning System
The presence of high concentrations of suspended solids can cause measurement errors when using UV sensors; long-term installation often necessitates frequent manual maintenance for cleaning sensor lenses. To address these issues, filtration of the original sample to reduce the concentration of suspended solids, combined with effective automatic cleaning, is essential for obtaining long-term stable and accurate measurements.
Introducing: SM57Q65 Automatic Filtration, Sampling, and Cleaning System
The SM57Q65 is an automated solution designed to prevent measurement interference caused by bubble accumulation, enabling stable long-term operation without manual intervention.
Operating Principle:
- Efficient Filtration: The ceramic membrane is installed in the water sample to be measured, and a peristaltic pump is used to continuously filter and sample the original sample.
- Precise Measurement: The UV device's measurement window is sealed within a specially designed flow cell to continuously measure the filtered sample.
- Automatic Cleaning: The cleaning system periodically injects an acidic cleaning solution into the flow cell to clean the measurement lens, flowcell, and sampling tube.
- Flexible Adjustment: The cleaning procedure and cleaning solution can be adjusted and replaced according to different application environments.
- Intelligent Control: When cleaning is in progress, measurement will stop, output will hold on.
Key Advantages:
- Reduced Maintenance: Stable long-term operation without frequent manual maintenance.
- Increased Accuracy: Filtration of suspended solids and elimination of bubble interference ensure accurate and reliable measurements.
- Flexible Adaptation: Customizable cleaning procedures and cleaning solutions to meet different application requirements.
- Intelligent Convenience: Automatic operation, easy to use, and reduced labor costs
-
-
Size259 (W) x 157 (H) x 147 (D) mm
-
Weight0.5KG
-
Maximum Sample Concentration35,000 mg/L
-
Post-Filtration Sample ConcentrationLess than 10 mg/L
-
Filtration MethodCeramic membrane
-
Post-Filtration Flow Rate0-200 mL/min, adjustable.
-
Maximum Sampling Height6m
-
Post-Filtration Sample ConcentrationContinuous sampling filtration with peristaltic pump.
-
Power supply24 VDC
-